笔者在日常检测中发现,滤波器通频带设置不当会造成心电图机幅值衰减、波形失真。尤其在检测“幅度时间参数测量”一项时测量值有可能超出规程限定值范围。因此,笔者认为“波形识别能力与幅度时间参数测量”一项也应当正确设置滤波器参数。一、低通滤波器设置对波形的影响1.设备与方法选择6台上海光电ECG-92C型数字心电图机(以下简称“心电图机”)。由EGC-1C心电信号发生器分别向6台心电图机输入频率为0.75Hz、幅度值为2mV的心电仿真信号,心电图机增益置为20mm/mV,走纸速度置为50mm/s。
The author found in routine testing, the filter pass band improper settings will cause the ECG amplitude attenuation, waveform distortion. Especially in the detection of “amplitude time parameter measurement” one of the measured values may exceed the regulatory limits. Therefore, I believe that “the waveform recognition capability and amplitude time parameter measurement” should also set the filter parameters correctly. First, the low-pass filter settings on the waveform 1. Equipment and methods to choose six sets of Shanghai Photoelectric ECG-92C digital ECG machine (hereinafter referred to as “ECG machine”). The EGC-1C ECG generator inputs the ECG simulation signal with the frequency of 0.75 Hz and the amplitude value of 2 mV respectively to the six electrocardiographs. The electrocardiograph gain is set to 20 mm / mV and the paper feeding speed is set to 50 mm / s.