我院自2001年7月引进美国VARIAN 2300C/D直线加速器,使用两年多来性能稳定,使用良好,但也出现过一些故障,其中以涉及真空泵、电离室电源盒的故障较多。下面介绍两例维修实例,供同行参考。 真空泵、电离室电源盒位于加速器机架右侧。真空电源盒内有高压、低压两块板和一只5000V变压器,高压板(VACION HV POWER SUPPLY)在电源盒的底层,外面有屏蔽箱,它的主要作用是为加速器提供5000V加速管主离子泵、
Our hospital since July 2001 the introduction of the United States VARIAN 2300C / D linear accelerator, the use of more than two years of stable performance, good use, but there have been some failures, including involving the vacuum pump, ion chamber power supply box failure more. Here are two examples of maintenance for peer reference. Vacuum pump, ionization chamber The power box is located on the right side of the accelerator frame. Vacuum power supply box has a high voltage and low voltage two plates and a 5000V transformer, high pressure plate (VACION HV POWER SUPPLY) in the bottom of the power box, a shielding box outside, its main role is to provide 5000V accelerator main ion pump ,