论文部分内容阅读
In this paper we present a timing detector which has low detection threshold and is radiation-resistant.The photons induced by radioactive beam ions passing through a thin plastic-scintillator foil BC422,emit from the foil center corresponding to one focal point of an aluminum ellipsoidal mirror and are reflected to another focus point at which the cathode of a photomultiplier tube locates.A time resolution of about 115 ps can be achieved for1 2N of 58.5 MeV/u and 130 ps for1 7B of 43.7 MeV/u.The counting rate up to 10 8packets per second(pps) is allowed.
In this paper we present a timing detector which has low detection threshold and is radiation-resistant. The photons induced by radioactive beam ions passing through a thin plastic-scintillator foil BC422, emit from the foil center corresponding to one focal point of an aluminum ellipsoidal mirror and are reflected to another focus point at which the cathode of a photomultiplier tube locates. A time resolution of about 115 ps can be achieved for 1 2N of 58.5 MeV / u and 130 ps for 7B1 43.7 MeV / u.The counting rate up to 10 8packets per second (pps) is allowed.