法国CRR SUPRA C 125移动式床旁X线机电路分析

来源 :医疗装备 | 被引量 : 0次 | 上传用户:cjl7648893
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
我院这台X线机是90年购买的,我负责维修保养多年。对它的电路进行了分析,现叙述如下供大家参考。一、高压电容充电a.检查球管不过热,即确定球管外的热接触器DJ101闭合;曝光前遮光器应闭合。其联动开关CT101闭合;曝光按键BP110二档断开;闭合主电源开关BP101;闭合钥匙开? This X-ray machine in our hospital was purchased in 1990. I was responsible for maintenance for many years. Its circuit is analyzed and is described below for your reference. First, high-voltage capacitor charging a. Check that the bulb is not overheated, ie, make sure that the thermal contactor DJ101 outside the bulb is closed; the shutter should be closed before exposure. The linkage switch CT101 is closed; the exposure button BP110 is opened in the second gear; the main power switch BP101 is closed; the key is closed?
其他文献
近年来,连云港产MG系列脉动真空灭菌柜以它的大容量、高效、带干燥等特点广泛应用于全国各大医院。现就本人在维修中的一些常见故障及维修方法介绍如下。故障现象1:真空键按下
故障现象:开机后,方式选择钮在辅助控制、同步间歇、自主呼吸位置,气道高、低压及自主呼吸指示灯全部报警;按报警消音按钮无效。同时,呼吸机各种呼吸功能丧失,但手控按钮尚可使呼吸
一台日本光电公司生产的三导心电图机ECG-53O3在开机使用时突然发现有故障,而前一天还在正常使用。其故障主要表现为导联不能切换,VI~V3各导均无信号。该机型是国内大中型医院比
故障现象:我院安装的该加速器在正常工作五年后,时常出现出束剂量到了预置数时不能自行停止现象,全靠操机人员监视控制台上的数码管指示按下“停止”键切断出束。因此误差很大,影
X光机是医院普遍装备的设备,为了不影响日常工作,X光机一旦出现故障就需马上修复。因此,常常需要采用应急维修措施,下面举例说明,如:一、F99一IAT型500mAX光机应急维修一例故障现象
COMPAK-200型准分子激光系统(MiniExcimerLascrSystem)是美国LASERSIGHT公司的产品.它用于眼科准分子激光角膜切削术(PRK手术)。此种手术治疗屈光不正有十分显著的疗效,是目前矫正视
超短波电疗机目前在各级医院使用较广泛,这种电疗机结构上较为简单,由以下两大部分组成。1.电路的高压部分及低压部份。2.高频振荡部分与输出部份。电源部份故障率极低,经常性出故
“江海”牌电动手术台是国产的几种电控手术台之一,该产品曾获江苏省科技进步二等奖和江苏省新产品“金牛奖”。然而在实际临床应用中却暴露出电气控制电路潜在故障率较高的缺
请下载后查看,本文暂不支持在线获取查看简介。 Please download to view, this article does not support online access to view profile.
期刊
杜仲(EucommiaulmoidesOliv.)有降低血压,补肾安胎的功效,主要成分为杜仲胶、氯原酸、桃叶珊瑚苷等。Charles从中国产杜仲中分离得到了松脂醇二葡萄糖苷(pinoresinddiglucoside),并确证了其为杜仲降压的有效成分[1 Eucommia ulmoides Ol