【摘 要】
:
故障现象:一台日本光电公司ECG-6511型心电图机,刚开机工作正常,5分钟后各个导联都无心电信号输出,定标信号也打不进。关掉仪器电源,等30分钟后,再开机工作又正常,几分钟后故障又出现
论文部分内容阅读
故障现象:一台日本光电公司ECG-6511型心电图机,刚开机工作正常,5分钟后各个导联都无心电信号输出,定标信号也打不进。关掉仪器电源,等30分钟后,再开机工作又正常,几分钟后故障又出现。心电图机面板指示一直是正常的。分析与检修:由于本机刚开机工作正常,几分钟后工
Symptoms: A Japanese photoelectric company ECG-6511 electrocardiograph, just started working properly, 5 minutes after each lead without ECG signal output, the calibration signal also can not enter. Turn off the power of the instrument, wait 30 minutes, then start the work again and normal, after a few minutes the fault appears again. The ECG machine panel indication has been normal. Analysis and maintenance: Since the machine just started working properly, after a few minutes