Testing and analysis on the protective performance of -knife being installed and used

来源 :首届全球华人辐射研究大会 | 被引量 : 0次 | 上传用户:xinmo2009
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
  Purpose: To grasp the radiation dose level from installation of γ-knife and its protective performance before it was used in medicine, and to improve the radiation monitoring system and quality control project with installation and application of γ-knife.Materials and methods: A head γ-knife typed OUR was used as the study object, other test tools such as SG-102 environmental X-γ dosimeter,451P radiation survey dosimeter, PTW UNIDOS dosimeter, chamber(0.125cm3 and 0.015cm3), φ 160cm PMMA spherical phantom and films(MD-55 and Kodak EDR2) were adopted.Results: (1) The dose rate measured from operation position, observation window were 45 μ Gy/h and 200 μ Gy/h respectively, while each 60Co medical source was being delivered from lead shielding container to the shielding body ofγ-knife, and the effective doses received from operators were (5.0-16) μ Sv in the course of sources installing(integral of 30min).(2) It is shown that the dose rate measured from the shielding surface of γ-knife, middle line of the surface of γ-knife therapy bodys shielding door, the location up and down of the the doors middle line were (0.35-21) μ Gy/h, 140 μ Gy/h, 79.7 μ Gy/h and 75.4 μ Gy/h respectively when the γ-knife was in the state of non-therapy.Simultaneously, the highest dose rate of the position 100cm away from the shielding body was 4.0 μ Gy/h in the same non-irradiation state.(3) The dose rate near thetreatment room wall facing the fnrst scatter field was (550-2570) μ Gy/h and the dose value outside this wall was (0.6-3.5) μ Gy/h in the radiotherapy with φ 18mm collimator, while the dose rate from the second scatter field were (5-30) μ Gy/h near the wall inside and (0.10-0.30) μ Gy/h near the wall outside in this treatment state.In addition, the radiation dose rate measured near the treatment room door were 12 μ Gy/h(inside) and 0.89 μ Gy/h (outside).(4) It is illustrated that the error between mechanical center and radiation field center of the γ-knife was 0.28mm, the focus dose rate was 351cGy/min(φ 18mm), the disparity between the projected dose and therapeutic dose was-1.0% and the ratio of focus radiation dose with "0"collimator versus "φ 18mm"collimator was 0.1%.Otherwise, relative helmet index(K) and penumbra of collimator(P) were measured, and they were described as follows: K(φ 18/ φ8)=1.000,K(φ14/φ18)=0.980,K(φ8/φ18)=0.897 and K(φ4/φ18) =0.570 while Pφ4, Pφ8, Pφ14 and Pφ18 were (2.0-3.2) mm, (2.8-5.2) mm, (3.1-7.3) mm and (3.7-8.9) mm respectively.(5) Basic safety and interlock were set to meet the clinic requirement and to avoid accidental irradiation.Conclusions: Firstly, The basic protective performance and safety should meet the relative requirements of γ-knife technical standard before.it was used in clinic.Secondly, all the performance should be tested in the normal application so as to assure the radiation safety and credible radiotherapy with the γ knife.
其他文献
随着市场经济的不断发展和完善,货运市场呈现出迅速发展的势头。如何提高运输经济效益?阜城县交通局运管站在广泛调查研究的基础上,针对县城车多、货多、地理位置优越的特点
西部情思 属马的人对马的钟爱似乎是出于本性,一匹棕色的伊犁马在古老的胡杨树下似睡非睡,树叶的投影在马背上凝聚成几片错落有致的花斑。也许这是一匹贯于劳作的普通马,或
1.CPU、显示卡等电脑配件不要过分地超频,要注意温度,否则,在启动或运行程序或是游戏的时候会莫名其妙地重启或死机.
转换企业经营机制,是当前经济工作的热点,受到各方面的高度重视,也是现实经济生活中长期没有解决的一个难点。从笔者掌握的部分水运企业情况看,机制转换中暴露的一些新问题
传统的数学教学过程只关注老师如何教、如何问,却忽略了学生如何学、如何想,这也导致学生课堂参与度不高,质疑能力和提问意识薄弱.实际上,小学数学正是夯实基础、培养兴趣的
会议
[摘 要] 阐述了圆锥曲线的定义及命题情况分析。  [关键词] 圆锥曲线 定义 命题
会议
智慧语文课堂不同于以前常规的语文课堂教学,它的优点是运用现代化的教学设备来使语文课堂变得充实,让小学生在课堂上可以用多种思维思考问题,充分体现智慧课堂,打造专属于小
会议