将SP-2305型气相色谱仪的分离系统,改装为二维气相色谱分离系统,采用二维柱切割、浓缩与转化技术相结合,以φ2×500mm,40—60目TDX-01柱为预分离柱;以φ3×2500mm,30—50目5A分子筛柱和φ4×1000mm,60—80目TDX-02柱为主分离柱。TCD和FID串联式连接,一次进样可测定高纯氩中O_2、N_2、CO、CH_4、CO_2、C_2H_2、C_2H_4、C_2H_6八种杂质。取10升气体样浓缩,方法灵敏度:O_2、N_2为0.1PPm,碳氧化物和烃类杂质为0.003———0.03PPm。
The separation system of SP-2305 gas chromatograph was converted into a two-dimensional gas chromatographic separation system, which was pre-separated by two-dimensional column cutting, concentration and transformation technology, and used φ2 × 500mm and 40-60 mesh TDX-01 columns Column; φ3 × 2500mm, 30-50 mesh 5A molecular sieve column and φ4 × 1000mm, 60-80 mesh TDX-02 column-based separation column. TCD and FID series connection, a sample can be measured high purity argon O_2, N_2, CO, CH_4, CO_2, C_2H_2, C_2H_4, C_2H_6 eight impurities. Take 10 liters of gas-like concentration, sensitivity: O_2, N_2 0.1PPm, carbon oxide and hydrocarbon impurities 0.003 --- 0.03PPm.